Correcting Faults; Technical Data - DENTAURUM 090-606-00 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6

Correcting faults

Error
Red "ERROR" indicator
on the casting machine
megapuls compact
lights up, audible
signal sounds.
Pump becomes noisy, air
bubbles can be seen in the
transparent hoses.
If you are not able to correct the fault using this table, please call your relevant service technician or call
Dentaurum Technical Service – Machines directly.
Tel. + 49 72 31 / 803 - 211 I info@dentaurum.com
7

Technical data

Mains voltage
Mains frequency
Output
Current
Electric fuse
Capacity
Pump output
Max. water pressure
Ambient air temperature for operation
Storage temperature
measurements [width x height x depth]
Net weight approx.
Reason
Water flow interrupted.
No electricity.
The water tank has not been mit filled
with enough water.
Air in the water circulation.
260 V
50 / 60 HZ
260 W
1.1 A
1.6 At
20 l
9 l / min
3.3 bar
15...40 °C / 59...104 °F
5...40 °C / 41...104 °F
680 x 650 x 600 mm
41 kg
Correction
Check hose connection between
megapuls compact and water
rotating unit.
Connect water rotating unit to the
socket on the reverse side of
megapuls compact.
Check fuse (1,6 At) at the reverse
side of the water rotating unit.
Check the water level and
add water if necessary.
Switch off the water circulation
unit and casting machine.
Unscrew the pressure hose with
angled connector ("WATER IN")
from the casting machine and hold
it in a container.
Turn on the casting machine until
water starts running out of the
hose. Switch off casting machine
and reconnect hose.
Switch on the casting machine
and check hose connections for
leakage. Add water if necessary.
EN
17

Publicité

Table des Matières
loading

Table des Matières