7. Cleaning and storage
Use an alcohol swab or cotton swab moistened with alcohol (70% Isopropyl) to
clean the thermometer casing and the measuring probe. Ensure that no liquid
enters the interior of the thermometer. Never use abrasive cleaning agents, thin-
ners or benzene for cleaning and never immerse the instrument in water or other
cleaning liquids. After cleaning the unit should not be used until the cleaning
agent is completely vaporized as the cleaning agent may affect the accuracy of
measurements. Take care not to scratch the surface of the LCD display.
8. technical specifications
Items
Model
Luvion Exact non-contact thermometer
Applicable
regulations
ASTM 1965 GB/T 19146-2010
and laws
Temperature
°C / °F, adjustable
units
Measurement
Body temperature mode: 32.0°C - 42.2°C / 89.6°F - 107.9°F
range
Milk temperature mode: 0.0°C - 100.0°C / 32.0°F - 212.0°F
Precision
±0.2°C / ±0.4°F
Display reso-
±0.1°C / ±0.1°F
lution
Latency Time
1 second
Abnormal
LCD Displays "L °C" if the measured temperature is below the
state display
minimum of measurement range.
LCD Displays "H °C" if the measured temperature is above the
maximum of measurement range.
10
Standards
LCD displays "Er1" if the temperature measurement circuit is
abnormal (fault of SENSOR or the temperature measurement
circuit), or the ambient temperature exceeds 10°C ~ 40°C
(50°F ~ 104°F). There shall be "ErC"if the calibration process is
not completed or EEPROM is abnormal.
Volume ≥50 db (the perpendicular distance from dB Volume
sensor to thermometer is 10cm).
Two kinds of backlit: red and green
Body temperature mode:
• There will be one long beep and green backlit for 3 sec-
onds when the temperature is between: 32.0°C/89.9°F and
Sound and
37.6°C/99.6°F
Backlit
• There will be seven short beeps and red backlit for 3 sec-
onds when the temperature is between 37.6°C/99.6°F and
42.2°C/107.9°F
Milk temperature mode:
• There will be one long beep and green backlit for 3 sec-
onds when the temperature is between: 0.0°C/32.0°F and
100.0°C/212.0°F
Automatic
shutdown
10s ±1s
function
Low-voltage
The product shall display low-voltage signal if the voltage is
display func-
below 2.61V ±0.15V
tion
Memory
Memorize 20 groups of measured temperature
function
Current
I
<2 µA; I
stand-by
working
consumption
LED backlit
Red ≥ 1.2cd/m
2
specifications
Green ≥ 2.0cd/m
2
Operational
ASTM 10°C - 40°C / 50°F - 104°F 15-95% RH
conditions
<0.5mA; I
<2mA; I
<15mA
Buzzer on
Backlit
11