• Don't use the non-contact thermometer in high humidity environments.
• Patients should not drink, eat, or be physically active before/while taking the
measurement.
• Before placing the thermometer sensor onto the measurement area, remove
hair, sweat, or dirt.
• Use an alcohol swab to carefully clean the sensor and wait for 5 minutes before
taking a measurement on another patient.
• 7 short beeps will sound when the temperature is higher than 37.5 °C in order
to alert the patient that he/she may have a fever.
• Always take the temperature in the same location, since the temperature read-
ings may vary from different locations.
• In the following situations it is recommended that three temperatures in the
same location be taken and the highest one taken as
the reading:
1) Newborn infants in the first 100 days.
2) Children under three years of age with a compromised immune system and for
whom the presence or absence of fever is critical.
3) When you're using the thermometer for the first time until you have familiar-
ized yourself with the instrument and obtain consistent readings.
6. Explanation of displays and operation instructions
LCD display
Operation instructions and explana-
tion of displays
1. Measurement of body
temperature
1. Aim the temperature sensor (1)
at the measurer's forehead directly
(effective distance 1 to 6 cm), under
shutdown state press button (2)
or hold down button (2), the LCD
display will display the measured
forehead temperature.
6
Sound and backlit
Body
temperature mode
1. When temperature
value is between
32.0°C and 37.6°C
there will be one long
beep and green backlit
for 3 seconds.
2. When temperature
value is
Holding down the button and scan-
ning over the forehead is suggested.
The system will read sequentially
and show the maximum value.
2. Measurement of milk tempera-
ture. Under shutdown state Press
button (3) and release it within 4
seconds. the LCD display will display
the measured milk temperature.
3. Press button (2) or (3) under
startup mode to measure again.
Attention:
There will be 7 short rapid beeps
when the measured temperature
exceed 37.5°C/99.5°F which is a
warning for patients that he/she
might have a fever.
The feeding bottle symbol will be
visible in the left top corner when
in Milk temperature mode. The °C
symbol will blink when the reading
or
is ready.
The forehead symbol will be visible
in the right top corner when in Body
temperature mode. The °C symbol
will blink when the reading is ready.
1. The measured value exceeds
100°C/212.0°F under milk tempera-
ture mode.
2. The measured value exceeds
42.2°C/107.9°F under body
temperature mode.
between 37.6°C/99.6°F
and 42.2°C/107.9°F,
there will be 7 short
beeps and red backlit
for 3 seconds.
Milk temperature
mode
1. When the tempera-
ture value is between
0.0°C/32.0°F and
100.0°C/212.0°F there
will be a long beep
and green backlit for 3
seconds.
3 short beeps and red
backlit for 3 seconds.
7