TROUBLESHOOTING
Switch off the device before any procedure and unplug the power cable from the socket.
PROBLEM
The device does not work
The device does not
nebulize or
nebulizes insufficiently.
The apparatus is noisier
than usual
If after verifying the above mentioned conditions the device should not operate properly,
contact your trusted retailer or the nearest authorized FLAEM service center.

ELECTROMAGNETIC COMPATIBILITY

This device was designed to satisfy the currently required requisites for electromagnetic
compatibility (EN 60601-1-2). Electro-medical devices require particular care during
installation and use relative to EMC requirements. Users are therefore requested to install
and/or use these devices following the manufacturer's specifications. There is a risk of
potential electromagnetic interference with other devices. RF mobile or portable radio and
telecommunications devices (mobile telephones or wireless connections) can interfere
with the functioning of electro-medical devices. For further information visit our website
www.flaemnuova.it. The Medical Device may be subject to electromagnetic interference
if other devices are used for specific diagnosis or treatments. Flaem reserves the right to
make technical and functional modifications to the product with no prior warning.
CAUSE
The power cable has not
been correctly inserted into
the socket of the device or
the power socket
The medication has not been
inserted in the nebulizer
The nebulizer has not been
properly fitted.
The nebulizer nozzle is
clogged
The air pipe is not correctly
connected to the device
The air pipe is bent, damaged
or twisted
The air filter is dirty
The Filter is not correctly
inserted in its housing
SOLUTION
Correctly insert the power cable
in sockets.
Pour the right amount of
medication in the nebulizer
Disassemble the nebulizer and
reassemble it correctly as shown
in the connection diagram on the
cover.
Disassemble the nebulizer, remove
the nozzle and perform cleaning
operations.
Failure to remove medication
deposits from the nebulizer affects
its efficiency and operation.
Strictly comply with the
instructions
contained in the "CLEANING,
SANITISATION, DISINFECTION"
chapter
Ensure the device's accessories
are properly connected to the air
intake (see connection diagram on
the cover).
Unwind the pipe and ensure it is
not crushed or punctured. If
necessary replace it.
Replace the filter
Fully and properly insert the Filter
in its housing
12