Vyaire Micro 1 Mode D'emploi page 43

Spiromètre de diagnostic
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 92
The Micro I has an essential performance – When verified with a 3 litre syringe
the readings remain within a tolerance of +/-3.5%, and unit firmware must not
cease operating. In the unlikely event of a Fast Transient / ESD event occurring,
the device should be reset and located away from the source of interference.
Note: In the very unlikely event of a high ESD event occurring to the device
connector, the device could be damaged.
WARNING:
The Micro I should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use with other equipment is necessary, the
Micro I and the other equipment should be observed / monitored, to verify
normal operation in the configuration in which it will be used.
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The Micro I is intended for use in the electromagnetic environment specified
below. The customer or the user of the Micro I should assure that it is used in
such an environment
Emission Test
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic
Emissions
IEC 61000-3-2
Voltage
fluctuations /
flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class [B]
not applicable
not applicable
Electromagnetic Environment
-Guidance
The Micro I uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The Micro I is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purpose.
41

Publicité

Table des Matières
loading

Produits Connexes pour Vyaire Micro 1

Table des Matières