1. Disconnect the air-hose.
2. Screw the air flow meter into the output thread of
the unit; then hold the unit in upright position at
eye-level.
3. Switch the filter unit on. The flow amount is satis-
factory if the flow meter float is in the green field.
If it is in the red field (see picture attachment), the
unit must be inspected (see Chapter 7).
5.
Maintainance and cleaning
• Cleaning must be done in a well ventilated room.
Beware of breathing in harmful dust settled on
individual parts of the filter unit and accessories!
• Do not use cleaning agents with solvents or
abrasive cleaners in any case.
• The outer surface of the filter unit can be cleaned
with soft fabric moistened in solution of water and
usual washing-up liquid. After cleaning dry indivi-
dual parts properly.
• Do not allow water or other liquid to get into the
filter unit!
• The air hose can be washed out with clean water
after disconnection from the unit.
6.

Spare parts and their replacement

6.1. Filters
CA CHEMICAL 2F
There are two filters attached to the unit with thread
RD40x1/7". The vital rule is to use a set of two
filters of the same type at a time !
Install only new filters without any damage.
It is forbidden to clean a filter or blow through it in
any way!
From the hygiene point of view it is recommended
not to leave a filter in the unit for more than 180
working hours.
It is also possible to use a pre-filter which holds
larger particles and especially aerosols which can
stick up the filter during paint-spraying and thus it
prolongs the life time of the filter.
REPLACEMENT OF FILTERS CA CHEMICAL 2F
see Picture appendix no.1
Picture 1A :
Demounting is done by unscrewing each filter sepa-
rately anticlockwise.
Caution! Before mounting new filters, make sure
the new filters are undamaged, in the original pac-
CA CHEMICAL - Filter selection in accordance with gas aplication (pollutant)
Supply number
Filter type *
50 00 48
P R SL
50 01 56
A
50 01 57
AP R SL
50 01 67
ABP R SL
50 01 68
ABEKP R SL
______________________________________________________________________________
king and their service life is not expired (expiry is
marked directly on the body of the filter). Check also
whether the seals at the connecting points of the
filters and the filter unit are unfaulted and safe.
Picture 1B :
Mount new filters one by one clockwise. It is ne-
cessary to be particular about screwing them tight
to ensure tightness of the joints.
CA CHEMICAL 3F
There are three filters attached to the unit with thre-
ad RD40x1/7". The vital rule is to use a set of
three filters of the same type at a time !
Install only new filters without any damage.
It is forbidden to clean a filter or blow through it in
any way!
From the hygiene point of view it is recommended
not to leave a filter in the unit for more than 180
working hours!
It is also possible to use a pre-filter which holds
larger particles and especially aerosols which can
stick up the filter during paint-spraying and thus it
prolongs the life time of the filter.
REPLACEMENT OF FILTERS CA CHEMICAL 3F
see Picture appendix no.2
Picture 2A :
Demounting is done by unscrewing each filter sepa-
rately anticlockwise.
Caution! Before mounting new filters, make sure
the new filters are undamaged, in the original pac-
king and their service life is not expired (expiry is
marked directly on the body of the filter). Check also
whether the seals at the connecting points of the
filters and the filter unit are unfaulted and safe.
Picture 2B :
New filters are mounted one by one clockwise. It is
necessary to be particular about
tight to ensure tightness of the joints.
Colour code
White
Solid and liquid particles (Aerosols)
Brown
Organic gases and fumes with boiling point >65° C
Brown
Organic gases and fumes with boiling point >65° C
White
Solid and liquid particles (Aerosols)
Brown
Organic gases and fumes with boiling point >65° C
Gray
Inorganic gases and fumes like Chlorine, H
White
Solid and liquid particles (Aerosols)
Brown
Organic gases and fumes with boiling point >65° C
Gray
Inorganic gases and fumes like Chlorine, H
5
screwing them
Main application
S, HCN
2
S, HCN,
2
Alwa