5 Error Codes

5.1 Error Codes
The Check O
Plus has self diagnostic features built into the software
2
to detect faulty readings outside of normal operating ranges. The
codes, descriptions and recommended actions are:
E01: Oxygen measurement out of range Hi ( ³ 102.0% calculated
by algorithm).
Recommended Action: Verify that the Check O
in the correct mode (Concentrator or Calibration Verification
mode). If error code repeats; perform a calibration verification. If
error code repeats again; Call Service Provider.
E02: Oxygen measurement out of range Low ( £ -2.0% calculated
by algorithm).
Recommended Action: Verify that the Check O
in the correct mode (Concentrator or Calibration Verification
mode). If error code repeats; perform a calibration verification. If
error code repeats again; Call Service Provider.
E03: Device memory corrupt or missing.
Recommended Action: Call Service Provider.
12
Plus is being used
2
Plus is being used
2
E04: Signal reading not stable.
Recommended Action: Call Service Provider.
E05: Pressure measurement out of Range Hi ( ³ 50 PSI).
Recommended Action: Check the pressure on a known gas source
pressure. If error code repeats; Call Service Provider.
E06: Outside of operating temperature Hi ( ³ 40° C).
Recommended action: The Check O
device closer to room temperature before use.
E07: Outside of operating temperature Low ( £ 15° C).
Recommended action: The Check O
device closer to room temperature before use.
E08: Device self check found error.
Recommended Action: Remove and replace the batteries. If error
code repeats; Call Service Provider.
Plus is too hot, cool the
2
Plus is too cold, warm the
2
1176119-A