8
(A) Power On
Press the "
/Set" button to turn on the display panel. Press the "
again to begin curing.
Press again to stop the device from emitting in the middle of a curing cycle.
Note: The unit will automatically snooze after 30 seconds of idle time to preserve battery power.
Note: The use of this device does not require any direct contact with patients.
Note: It is recommended to use disposable sleeves before operating the device.
Note:This device is only to be used by trained personnel in dental practices.
WARNING :
When the device is in use, do not look directly into the light beam. Looking
directly into the light beam for an extended duration will cause eye damage.
CAUTION :
Avoid direct contact of light guide with skin, oral tissue and tooth surface. Before
use, ensure to put anti-glare cone on the tip of the light guide and keep the light
guide at a distance of minimum 2mm from oral tissue or tooth surface.
CAUTION :
It is forbidden for the Fiber Optic Light Guide to be directly in contact with any
of the teeth surface or any other oral tissue.
CAUTION :
After each curing cycle, it is necessary to cool down for more than 30 seconds
before starting the next cycle.
Any clinical diagnosis need to be enforced by doctor according to the curing
environment and each patients' situation.
(B) Select the Mode
Press the "Mode(▲ " button to select the mode desired.
M1: Full Power Mode - Full power light intensity.
M2: Ramp-Up Mode - Light intensity gradually climbs from low to high.
M3: Pulse Mode - Alternates between strong and weak light intensity every
half second.
Note: When replacing or reinstalling the battery, the default operation mode of the device
is "M1".
Note: When the device snoozes, it will restart with the last mode selected.
(C) Select the Curing Time
Press the "Time(▼ button to select the curing time desired.
M1: Full Power Mode(1, 2, 3 secs)
M2: Ramp-Up Mode(5, 10 secs)
M3: Pulse Mode (10, 20, 30 secs)
Note: Default curing cycle for M1 is 3 seconds, M2 is 5 seconds, and M3 is 10 seconds.
Note: After replacing the battery, all user operation settings will revert back to the factory settings.
Note: To avoid overheating, it is recommended to avoid overextending the cycle duration.
Panel Operations:
)
)"
/Set" button