Calibration Oft He Scale; Troubleshooting - Wohler DTEST Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Digital Mini Scale
3.8
Calibration oft he
scale
3.9

Troubleshooting

28
Turn the scale on and wait fort he display to
show „0.000 g".
Press and hold the MODE key until the display
shows „CAL". Then release.
Press the MODE key again
The display will flash „50.000 g".
Place the calibration weight on the center of
the platform.
The display will show „PASS", then return to normal
weighting mode.
The primary reasons for inaccuracy or malfunction
are low batteries, incorrect calibration, overload or
operating on an unstable surface. The following
error messages can appear in the display
Error message
LO (Low voltage)
OUR 2
O-Ld
UnSr
Solution
Change batteries
Calibrate scale
Overload. Remove the
excessive load imme-
diately
Use the scale in a
more stable position