12. Click to exit the program
- Click the "START" from main screen to download the memory from logger
SpECIFICATIoNS
Description
Temperature range
Humidity range
Accuracy:
Temperature
Relative Humidity
Sample points
Resolution:
Temperature
Relative Humidity
LeD response
LCD Display
Dimension
Housing
TRoUBLESHooTINg
Cannot connect and the display won't turn on
Ans: Make sure you have selected the correct baud rate. Select 9600 baud rate for all models.
Meter is programmed as Magnetic but could not start
Ans: Set up Magnet start in the Download Suite software. Use a magnet on the rear side of
the meter for more than 2 seconds, the logger should be activated and the ReC flashes. if not,
change a new and stronger magnet and try it again.
TR100-A
-40 to 85°C(-40 to 185°f)
n/A
± 0.6°C / 1°f (-20 to 50°C / -4 to 122°f),
±1.2°C / 1.2°f all others
n/A
1K / 2K / 4K / 8K / 12k / 16K
0.1°C / 0.1°f
n/A
Red: Hi, Lo Alarm; Yellow: Record
13 x 13 mm
75 x 55 x 23 mm (2.95 x 2.16 x 0.9 in)
ip65
9
TR200-A
0 to 100 %RH
±3% (25°C, 10-90%),
±5% all others
0.1%
ip54