Télécharger Imprimer la page

Mettler Toledo LC-RS25 Mode D'emploi page 2

Câble

Publicité

Matching the cable to peripherals
The cable is configured in the factory for attachment of a computer (printed in bold).
The following protocol must be set at the peripheral device: 2400 baud, 7 bits, even parity, CTS/DTR.
If other devices need to be attached or work is performed with a different communications protocol, the cable must be
appropriately configured using the 3 switches.
Switch left
Pos.
Device
0
Host (PC)
Computer, database
1
Printer (24)
Strip printer, 24 characters/line
2
Printer (80)
Page printer A4, 80 characters/line
3
2nd balance Second balance, reference balance
4
ID unit
Barcode reader, terminal
5
Gen. I/O
I/O module
6
Spec.
I/O special applications
7
Alt.
other devices, older METTLER TOLEDO
devices; in this configuration,
special settings apply to the
right switch; for operation with
8 bits, only baud rate 9600
is possible.
Switch middle
Switch right
Pos.
Baud rate
Pos.
0
300
0
1
600
1
2
1200
2
3
2400
3
4
4800
4
5
9600
5
6
19200
6
7
unassigned
7
0...6
300...
0
19200
1
2
3
4
5
9600 (fixed)
6
9600 (fixed)
7
Protocol
7 Bit, Even, CTS/DTR
7 Bit, Odd, CTS/DTR
7 Bit, Mark, CTS/DTR
8 Bit, NoPar, CTS/DTR
7 Bit, Even, Xon/Xoff
7 Bit, Odd, Xon/Xoff
7 Bit, Mark, Xon/Xoff
8 Bit, NoPar, Xon/Xoff
7 Bit, Even, Syn-Ack
CL handshake
7 Bit, Even, CTS/DTR
CL free mode
7 Bit, Even, Pause 1 s GA44 protocol
7 Bit, Even, CTS/DTR
T, S, SI, SIR
7 Bit, Even, CTS/DTR
Continuous (T, S, SI, SIR)
7 Bit, Even, CTS/DTR
Option 03
8 Bit, No, CTS/DTR
Auxiliary display
8 Bit, No, Xon/Xoff
Reserved

Publicité