OPERATION INSTRUCTIONS
04 Transfer now
04-Transferring
0
05 Clear memory
05-Clear mem?
No
Yes
▼▲ to change
Enter
04SCALE SETTINGS
Main Menu
02-Tare Weight
03– Data Management
→
04– Scale Settings
▼▲ to change
05-System Settings
06– System Test
07– User Settings
08-Exit
Before making any changes to this scales settings, please refer to the Menu instructions on page 7 to
temporarily activate the Menu key.
01 Auto Hold Time
01-Auto Hold Time
00
Seconds
Enter
All the values stored in memory are transferred to the
computer and the scale memory is cleared of all values. If
10
the transfer was unsuccessful, the values are kept in
memory until successfully transfer or cleared. Press EX-
IT to return to the previous screen.
If Yes is selected and Enter is pressed, all the values in
memory will be cleared. If No is selected, the display will
return to the last menu. Press EXIT to return to the pre-
vious screen.
▲
▼
Enter
The user can determine how long to display the weight reading once
it is determined, regardless of whether the patient remains on the
platform. The scale defaults to no Auto Hold Time. The maximum
setting is 20 seconds Hold Time. To set the time, use the keypad to
enter the desired number of seconds, and then press the ENTER
key. The display will then return to the Scale Settings menu.
04-Scale Settings
→
01-Auto Hold
02-Sleep Time
03-Auto Off Time
▼▲ to change
04-Tone Volume
05-Disp Date & Time
06-Display Backlight
07-Display Contrast
08-Live Weight
09-Exit
12
▲
▼