Health O Meter PROPLUS Mode D'emploi page 11

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

OPERATION INSTRUCTIONS
Removing the Tare
Once a tare value has been stored and the word ―T A RE‖ appears on the display, the tare value can be deleted from
memory by pressing the TARE button. The scale will return to normal operation.
03 DATA MANAGEMENT
Before making changes to this scale setting, please refer to the Menu instructions on page 9 to temporarily activate
the Menu key
The scale managed patient data including patient ID, weight, height, tare, and BMI. The tare is stored in memory or
transferred to a PC. This function is performed by opening a new data file. The maximum capacity of the scale is
270 files of different data.
1.
Press the ID button (7) See page 4 for the illustration
2.
Using the keypad, type in the identification number
3.
Press ENTER
01 Auto-Download
01-Auto Download
Updated
02 Store in memory
02-Store In Memory
Updated
03 Do not store data
03-Do not store data
Updated
03-Data Management
01-Auto Download
02-Store In memory
03-Do not store data
▼▲ to change
Enter
04-Transfer now
05-Clear memory
Automatic Download is the default option and will transfer the
value to the computer as soon as the patient steps off the
scale or when the user presses the HOLD/RELEASE button if
it was kept in ―HO LD‖. If no computer is connected, the value
is not transferred or stored and will be lost after the load is re-
moved from the scale.
The value is stored in memory for
later download to a computer. If the
memory is full the user will be
warned and given the option to
transfer all values to the computer
or to clear the memory of values.
Press EXIT to return to the previous
screen.
No data will be stored.
Press EXIT to return to the
previous screen.
11
Memory FULL
8 positions left in memory
01-Transfer now
02-Clear memory
▼▲ to change
Enter

Publicité

Table des Matières
loading

Produits Connexes pour Health O Meter PROPLUS

Table des Matières