Operating instructions
T
S
Figure 1: Views on Ventrain
= thumb hole,
T
1
Pre-oxygenate the patient (if possible)
2
When no suitable ventilation lumen is intubated, insert a catheter
according to its user instructions.
3
Connect the oxygen tubing of Ventrain
4
Connect the capnometer to the side port next to the catheter
connection or, when using no capnometer, make sure the cap is
tightened.
5
Open the oxygen supply and set a relatively low initial flow (rule
of thumb: start with 1L/min per year of age with a minimum of
2 L/min and a maximum of 15 L/min). In case of ventilation of a
collapsed adult lung use an initial flow of 4 L/min.
6
Hold Ventrain
®
no oxygen is flowing into the lungs and almost no gas is sucked
out.
®
= index finger hole
S
in the equilibration position as shown in Figure 2:
T
S
to the oxygen supply.
®