Hach Polymetron NA9600 sc Instructions D'utilisation page 19

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 55
Option
TEST OUTPUTS
STATUS
MODULE
INFORMATION
DEFAULT SETUP
Section 7 Troubleshooting
For general problems with HART module communications, try the corrective actions listed below. If
the problem continues or other problems occur, contact technical support.
• Make sure the connections between the module and the HART host are correct.
• Make sure the loop supply is live.
• Make sure the HART host is set to the correct device address.
• Check the module network setup for the selected channel. Make sure that the installed device is
mapped to the selected channel. Make sure the SET LOW VALUE and the SET HIGH VALUE are
within the range of the selected source.
• If the HART output from the HART host cannot be trimmed or fixed, make sure that the analyzer
does not have an error that makes the analyzer hold or transfer the output at a fixed value.
Description
Sets a value (mA) on the selected output. The simulation stops after the
user exits the screen.
Displays the values at the outputs (4–20 mA).
Displays information about the installed HART module.
• Software version
• Bootloader version
• Serial Number
• Driver version
• HART version
Sets the configuration to factory defaults.
English 19

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Hart module

Table des Matières