Hach Polymetron NA9600 sc Instructions D'utilisation page 16

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 55
Figure 11 HART voltage circuit - active module powered
1 HART module
2 Hand-held terminal (Secondary
communication device)
Figure 12 HART current circuit - active module powered
1 HART module
2 Hand-held terminal (Secondary
communication device)
Table 4 Resistance values for active module powered HART circuits
Loop resistance
250–350 Ω
4.12 Circuit diagrams for non-HART communication
Figure
13,
Table
5,
Figure 14
Figure 13 4-20 mA circuit - active module powered
1 HART module
Table 5 Resistance values for active module powered 4-20 mA circuit
Power supply voltage
15 VDC
16 English
and
Table 6
are not recommended for HART communication.
3 External device (Recorder/data acquisition
system)
3 External device (Recorder/data acquisition
system)
2 External device (Recorder/data acquisition
system)
Loop resistance
0–350 Ω

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Hart module

Table des Matières