Malfunctions • Actions; Connection To A Media Supply; Connection In The Media Duct; Connection In The Media Socket - asecos GAP Série Notice D'utilisation

Postes de travail destiné aux matières dangereuses
Masquer les pouces Voir aussi pour GAP Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

4. MALFUNCTIONS • ACTIONS
a
b
c
d

5. CONNECTION TO A MEDIA SUPPLY

5.1. CONNECTION IN THE MEDIA DUCT

BELOW WORKTOP

5.2. CONNECTION IN THE MEDIA SOCKET

ON BACK PANEL
Malfunction
• the mains plug is inserted
• a does not illuminate
• the On/O switch had been activated
• b does not illuminate
• the light switch had been activated
• The integrated lighting doesn't go on
• c illuminates during operation of the workplace
• 3 short beeps sound every 60 seconds
• c shortly lluminates every 20 seconds
CAUTION:
If your hazardous material workplace has the above or other defects, then please contact your dealer.
4×20
TX20
1
G 1/2
4
1
• Check if there is power on the socket in use
• LED defective (replace) or
• defect in the electronics
• Replace the light bulb.
• Please check the pressure loss on customer's exhaust
air duct (see these instructions hazardous material
workplace! under ventilation adjustment/monitoring of
the hazardous material workplace) or
• defect of the integrated air inlet fan or
• heavy soiling of the discharge rails.
• power failure
20 cm
2
2
action
3
G 1/2
3
EN
13
13