6. CALIBRATION
Press
SET
It displays
E
*
It displays
P
*
It displays Max
capacity like
05000
It displays
noload
It displays
F.S.
Finished
7. CHANNEL CONFIGURATION
1.
Press the
key, the display will show P00 , enter the code 20 and press
SET
2.
The display will show pd 01 . 01 represents the current channel number. Enter the desired channel number and press the
key, the display will show END, press
Remember that once a new indicator is paired, you must proceed to calibrate the crane scale.
8. WARRANTY
This scale is warranted against defects of manufacturing and materials for a period of 1 year, from the delivery date.
During this period, GIROPÈS SL will take charge of repairing the scale
This warranty does not cover defects or damaged caused by misuse, overloading or improper installation contrary to the
recommendations described in this manual.
This warranty does not cover shipping costs for the reparation of the balance.
Input password 13386 and press
means division value, you can choose
*
* means decimal points of division value, choose
press
input Max Cap you want and press
Wait until stable light on and press
Load weights onto the scale (like 3000kg), input value of weights like 03000, wait until the stable light on,
then press
again to confirm and finish.
and then press
1, 2, 5, 10, 20, 50
for no decimal or
0
.
USER'S GUIDE IMP20
means 1 decimal; then
1
EN
EN
21
21