5. PARAMETER SETTING
5.1 ESSENTIAL PARAMETER SETTING
Press
and input
18
SET
F1 1
F2 1
F3 05
F4 4
F5 4
5.2 SETUP PRINTING PARAMETER
Press key
, input
SET
P1 0
P 2 1
P 3 0
20
20
, then press
Zero function range setup, choose 0 means 0%, 1 means 2%, 2 means 20%, 3 means 50%,
4 means 100%, then press
Zero function range of power on, choose 0 means 0%, 1 means 4%, 2 means 50%,
3 means 100%, then press
Zero-tracking range, choose 5 means 0.5e, 10 means 1e, max is 20 means 2e,
then press
to confirm.
Baud rate of wireless, choose 1 means 600bps, 2 means 1200bps, 3 means 4800bps,
4 means 9600bps, then press
Baud rate of RS232, choose 1 means 600bps, 2 means 1200bps, 3 means 4800bps, 4 means
9600bps, then press
and press
97
Setup print out whether back to 0 or not, choose 0 means back to 0 and print, 1 for not back
to 0 and print, then press
Setup copies of printing bill, choose 1 means 1copy, 2 means 2 copies, 3 means 3 copies, then
press
to confirm.
Setup time of auto printing after stable light on, choose 0 means the time after stable, 1 means
1s after stable, 2 means 2s and the longs time is 3s.
to confirm.
to confirm.
to confirm.
to confirm.
to confirm.
USER'S GUIDE IMP20
EN
EN