For each individual patient it's recommended to use the nebulizer for 6 months or for a maximum of 120
treatments. The nebulizer must be replaced after a long period of inactivity, if it is deformed or broken, or if
the nebulizer nozzle is blocked by dry medicine, dust, ecc..
Only use the original nebulizer supplied by CA-MI with the device
The air filter must be replaced every 25 hours of functioning or when it result particulary worn.
For replacement, lift the filter and replace with a new one.
Only use original CA-MI filter
DON'T USE THE DEVICE WITHOUT AIR FILTER
The MIKO atomiser does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use. Unpack the instrument and always
check integrity of plastic parts and feeding cable., they might have been damaged during previous use.
Connect cable to electrical network and turn switch on.
Close the compressor mouthpiece with a finger and verify that loud noises are not present, these can indicate
wrong functioning. Verify that the atomiser is not damaged by previous use (it was badly put away or badly
knocked). A protection fuse (F 1.6 A 250V) not reachable from exterior protects the instrument.
For fuse replacing, please make reference to manufacturer technical personnel.
Fault type
1. Low Nebulization
2. Low Nebulization
3. Absence of Nebulization
4. Slow Nebulization
5. Noisy Device
Fault 1 - 2 - 3 – 4 – 5
If the unit does not nebulizer once the above conditions have been checked, we suggest to contact your
dealer or technical service CA-MI.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICL SERVICE

STANDARD ACCESSORIES

ACCESSORIES
HI-FLO KIT
(Nebulizer HI-FLO, Adult Mask, Pediatric Mask,
Air Tube and Mouth-piece)
Air Filter (+n°1 spare part)
Carrying Bag (on request)
MAINTENANCE
Cause
Clogged Nebulizer Tank
Clogged Nebulizer Tank
Clogged Nebulizer Tank
Highly drug
Extended use
No solution with previous
items
PLEASE CONTACT CA-MI TECHNICAL SERVICE
CHECKING, APPEARS TO BE TAMPERED
Clean and disinfect the nebulizer tank as
explained in the instruction manual
If cleaning was not succesful change cruet
Check that the nebulizer contains
medication; Make sure that the nebulizer is
Check the connection between the
compressor air outlet port and the
Dilute drug in physiological liquid
Call retainer or manufacturer CA-MI
Call retainer or manufacturer CA-MI