Guidance and manufacturer's declaration – Electromagnetic Emissions
The MIKO aerosol is intended for use in the electromagnetic environment specified below.
The customers or the user of the MIKO aerosol should assure that it's used in such an environment.
Emissions Test
Irradiated / Conducted emissions
CISPR11
Irradiated / Conducted emissions
CISPR11
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations /
flicker emissions
IEC/EN 61000-3-3
Guidance and manufacturer's declaration – Electromagnetic Emissions
The MIKO aerosol is intended for use in the electromagnetic environment specified below.
The customers or the user of the MIKO aerosol should assure that it's used in such an environment.
Immunity Test
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
Electrical fast transient / burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Loss of voltage, brief voltage
interruptions and variations
IEC/EN 61000-4-11
Magnetic field
IEC/EN 61000-4-8
Conducted Immunity
IEC/EN 61000-4-6
Irradiated Conducted
IEC/EN 61000-4-3
Note U
is the value of the power supply voltage
T
(as request by regulation EN 60601-1-2:2001)
Compliance
Group 1
Class [B]
Class [A]
Complies
(as request by regulation EN 60601-1-2:2001)
Compliance
± 6kV on contact
± 8kV in air
± 2kV power supply
± 1kV differential mode
5%U
for 0.5 cycle
T
40%U
for 05 cycle
T
70%U
for 25 cycle
T
<5%U
for 5 sec
3A/m
3Vrms 150kHz to 80MHz
(for appliances that aren't life -
supporting)
3V/m 80MHz to 2.5 GHz
(for appliances that aren't life -
equipment)
14
Electromagnetic environment -
The MIKO aerosol only used RF
energy only for its internal functioning.
Therefore its RF emissions are very low
and are not cause interference in
proximity of any Electronic appliances.
The MIKO aerosol can be used in all
environments, including domestic and
those connected directly to the public
mains distribution that supplies power
to environments used for domestic
scopes.
Electromagnetic environments -
Floors should be wood, conceret or
ceramic tile. If floors are coverei with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a
typical commercial environment or
hospital
Mains power quality should be that of a
typical commercial environment or
hospital
Mains power quality should be that of a
typical commercial environment or
hospital If the user of the MIKO aerosol
request that the appliance operates
T
continuosly, the use of a continuity unit
is recommended.
The power frequency magnetic field
should be measured in the intended
installation location to assure that it's
sufficiently low.
guidance