Pulmonetic Systems LTV 1200 Série Manuel D'utilisation page 277

Table des Matières

Publicité

Nom
événeme
nt
Code
USAG
STACK
56
POST
57
ER
58
CODE
59
WDOG
TST
60
CLR CAL
61
XDCR
NAR
62
XDCR
WID
63
XDCR BI
64
XDCR
AIR
65
ADC1
VAL
66
TBN
HSTP
67
LN
VENT1
68
FLUSH
ER
69
RAC
ERR1
70
RAC
ERR0
71
SNDRER
R1
LTV 1200™ Series Ventilators
Evénement
Débordement de batterie détecté
Echec POST
Code invalide détecté
Watchdog test run
Calibration records cleared
Differential pressure transducer - Narrow
channel fault
Differential pressure transducer - Wide channel
fault
Differential pressure transducer - Bi-directional
channel fault
Airway pressure transducer fault
AD mismatch primary channel fault value
Turbine Hold Stop occurred
Shutdown for other than pressing On/Standby
button
A problem is detected writing data to the
EEPROM during system shutdown.
Problem detected with primary and/or
redundant audible alarm circuitry
Recovery from problem detected with primary
and/or redundant audible alarm circuitry
Alarm sounder error
Alarme correspondante
RESET
RESET
RESET
Inop
N/A
XDC FLT1
XDC FLT1
XDC FLT1
XDC FLT1
HW FAULT
HIGH PRES
RESET
HW FAULT
HW FAULT
HW FAULT
HW FAULT
Page E-5