How to adjust an FK Force Measuring Device:
1. A short warming up time of approx. 1 minute is
recommended for stabilisation.
Switch on the FK by pressing "ON/OFF" button while
holding it in vertical position.
2. When display shows zero, simultaneously press "UNIT"
and "ZERO" button until "CAL" appears on display.
3. Wait a few seconds and display will show the necessary
adjustment weight, e.g. 5,000kg for FK 50 or 10,00kg for
FK 100.
4. Put the corresponding weight at the loading hook. Avoid
trembling and keep still while holding the weight (hint:
hold device with both hands while shoring up your elbows
on table or desk).
FK-BA-f-1213
Sauter GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@sauter.eu
Mode d'emploi
FK
5. Hold position until "F" appears on display. In this case
remove weight and adjusting is completed. In case of "E"
showing up on display - repeat step 1. to 5. until "F"
attests a successful adjustment
Tel: +49-[0]7433- 9933-199
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
4