Télécharger Imprimer la page

Tintometer Lovibond MD50 Manuel page 47

Masquer les pouces Voir aussi pour Lovibond MD50:

Publicité

Chlorine HR 2 PP / M112
Chemical Method
DPD
Appendix
Interferences
EN
Persistant Interferences
• All oxidising agents in the samples react like chlorine, which leads to higher results.
Removeable Interferences
• Interference from Copper and Iron (III) are eliminated by the addition of EDTA.
• Concentrations above 10 mg/L Chlorine, in the event of using Powder Packs, can lead
to results within the measuring range of up to 0 mg/L. In this case, the sample must be
diluted with chlorine-free water. 5 ml of the diluted sample should be mixed with the
reagent and the measurement taken again (plausibility test).
Conformity
EN ISO 7393-2
48
EN   Method Reference Book   1.0

Publicité

loading

Ce manuel est également adapté pour:

Lovibond md150