Dissolve tablet(s) by
inverting.
2 min
12
11
1
10
2
9
3
8
4
7
5
6
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L total Chlorine appears on the display.
Determination of Chlorine differentiated with tablet
Select the method on the device.
In addition, choose the test: differentiated
For this method, a ZERO measurement does not have to be carried out every time on
the following devices: XD 7000, XD 7500
10 mL
Fill 24 mm vial with 10 mL
sample.
EN Method Reference Book 1.0
Sample
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Close vial(s).
Chlorine T / M100
Test
Press the TEST (XD:
START)button.
Sample
Place sample vial in the
sample chamber. Pay
attention to the positioning.
EN
11