Removal / Transfer Option 1:
1. Turn off suction. Disconnect tubes from FILTRON lid, maintaining sterility.
2. On lid, close VAC port with attached cap. All ports except PATIENT should now be closed.
3. Attach the Transfer Tube to PATIENT transfer port. See Figure 5.
Close the white pinch clamp on the Transfer Tube. All ports are now closed.
4. Carefully invert the FILTRON upside down.
5. Reattach inverted canister to the canister stand using the same bracket.
6. Attach Luer Lock syringe to Transfer Tube. Open pinch clamp and draw out fat to be re-injected. See Figure 6.
Repeat with additional syringes.
7. When desired volume has been transferred, discard entire system.
8. Attach re-injection cannula. Fat is ready to be re-injected to recipient site (if applicable).
Removal / Transfer Option 2:
1. Turn off suction. Disconnect tubes from FILTRON lid, maintaining sterility.
2. On lid, close both VAC and PATIENT ports with attached caps. All ports should now be closed.
3. Attach transfer or re-injection cannula to luer lock syringe.
4. Remove cap from large CAP1 Port.
5. Insert cannula into CAP1 Port and draw out fat to be re-injected. Repeat with additional syringes.
6. When desired volume has been transferred, discard entire system.
7. Attach re-injection cannula. Fat is ready to be re-injected to recipient site.
Dispose in standard medical biohazard waste receptacle.
Device is inverted.
Device remains upright.
Do Not Re-sterilize or Re-use.