The AC charging adapter supplied with Oska Pulse is a certified Medical Grade
adaptor (Globtek, type GTM46101-1005-USB). Contact Oska, Inc. or local distributor
to obtain a replacement. If using another AC adaptor other than Oska supplied, please ensure a
certified adaptor to medical device safety standard (IEC 60601-1) with an output marked Class 2 or
meeting limited power source (LPS) requirements, and not to exceed 16V.
When the device is connected to a power source, LED lights will illuminate on the opposite
end. If the battery is fully depleted then the LEDs may take a few minutes to illuminate.
LEDs will remain lit (not pulse) while the device is charging, then go off when charging is
complete. The battery may take three (3) or more hours to fully charge. Battery charge
time may vary by environmental conditions.
A fully charged battery can provide ten (10) or more 90-minute sessions. Device run time
can vary due to operating conditions.
Battery life may decline following extended use. The battery is estimated to last between
200 and 500 charge cycles, but life may vary. Storing the device uncharged can reduce
battery life. The battery is not user replaceable.