OPERATING MODES
(Specifications apply for the top decade of each frequency range and output 10V peak-to-peak
into 500 termination).
SINE
Distortion:
Amplitude Flatness:
TRIANGLE
Linearity:
SQUAREWAVE
Rise and Fall Times:
Mark-Space Ratio:
DC
Range:
SYMMETRY
Symmetry Range:
OUTPUTS
500
0dB Range:
-20dB Range:
DC Offset Control
Range:
6004 (not TG230):
QUE
SWEEP OUT
(TG230 only)
GENERAL
POWER REQUIREMENTS
AC Input Voltage:
Power Consumption:
Operating Range:
Storage Range:
Environmental:
Size:
Weight:
Safety:
EMC:
Less than 0.5% on 200, 2k and 20k ranges; less than 1% on 2, 20 and
200k ranges; all harmonics >25dB below fundamental on 2M range.
+0.2dB to 200kHz; +1dB to 2MHz.
Better than 99% to 200kHz.
<100ns.
1:1 +1% to 100kHz.
+10V (unterminated)
Variable typically between 1:9 and 9:1 (on top decade of each range),
frequency divided by 10. Display shows correct frequency.
Two switch-selectable ranges with 40dB vernier control within each
attenuator range.
0.2V to 20V peak-to-peak from 50Q (0.1V to 10V into 500).
20mV to 2V peak-to-peak (10mV to 1V into 500).
+10V from 50Q. DC offset plus signal peak limited to +10V (+5V into 500).
DC offset plus waveform attenuated proportionally in -20dB position.
Alternative output socket offering the same facilities as the 50Q socket.
Capable of driving CMOS and 2 standard TTL loads.
6V ramp from 6000.
220V-240V AC or 110V-120VAC +10%, 50/60Hz,
by rear panel adjustment; Installation Category II.
25VA max.
+5°C to +40°C, 20% to 80% RH.
-10°C to +65°C
Indoor use at altitudes up to 2000m, Pollution Degree 2.
260(W) x 88(H) x 235(D) mm
(10.2 x 3.4 x 9.2") excluding handle and feet.
1.9kg (4.21b).
Complies with EN61010-1.
Complies with EN61326.