WARNING
• Do not attempt to measure more than 2000µA.
Features:
DC Microamps Measurement -Test lead input
Range
600µA
0.1µA
2000µA
1µA
AC Microamps Measurement -Test lead input
Range
600.0µA
0.1µA
2000µA
1µA
45Hz to 400Hz True RMS
The new DL589 clamp meter offers a digital field temperature calibration
procedure to make the process easier. Here are the steps.
1) Set dial position to "°F°C" and check T1 mode
2) Press and hold SELECT button for about 2 seconds
3) Press HOLD button while "FCAL" is displayed on top of the LCD to enter
Field Calibration.
4) After immersing T1 temperature probe in ice bath and the temperature is
stable, press and hold Hz/Duty button
5) Press SELECT button to switch to T2 mode.
6) After immersing T2 temperature probe in ice bath and temperature is
stable, press and hold Hz/Duty button to complete Field Calibration
AC/DC Microamps: <2000µA
CAT IV
600V
600A
DL589
A
Temp
µA
V
SELECT
Low Z
A
AUTO
V
A
PEAK
True RMS
30V
T
T
2
1
TEMP
SELECT
BT
Resolution
Resolution
Temperature Calibration
• Default =
• Press SELECT =
• Readings show on lower display
Accuracy
±1.2% + 3dgts
Accuracy
±1.5% + 3dgts
11
Overload Protection
600V RMS
Overload Protection
600V RMS