Télécharger Imprimer la page

Maintenance And Troubleshooting - Mesis JoySense 2.0 Mode D'emploi

Publicité

Les langues disponibles
  • FR

Les langues disponibles

User Manual: JoySense

Maintenance and Troubleshooting

The accessories cannot be repaired, as such take care of them.
Do not place the leggings, armband, abdominal belt or other accessories
close to direct or indirect sources of heat.
Do not place the equipment in damp or wet areas. Do not soak the
equipment in any way, do not pour any liquid over the equipment.
Do not wash the equipment and accessories with water or detergents.
Simply wipe with a damp cloth.
Do not place the equipment and its accessories in very cold areas or
areas below zero degrees centigrade. It could cause serious damage.
If you don't use the device for a length of time, place it back in its box.
Environment of use and safekeeping
• Use and maintain the device and its accessories in a clean environment and at a
temperature between 10 – 40 °C, and humidity of 30 – 85%;
• Do not use outdoors.
Problem found
The machine does not
turn on
If you hear a strange his-
sing sound similar to an
air outlet
The air does not flow to
the applicators, there is
insufficient intensity
The air continues to flow
into the chambers and
you hear an air leak in
the applicators
2.0 e JoySense
3.0
®
®
Possible solution
• Check that it is correctly connected to the mains;
• Check that you have inserted the air tube connector to the machi-
ne;
• Check the condition of the power cord. It must be intact.
• Check that the tubes, accessories and connectors are intact;
• Check that all the connections (to the machine and the applicators)
have been correctly inserted;
• Check that the air connection tubes are not bent, squashed or
crushed.
• Check that the connectors to the machine and the applicators are
correctly connected;
• Check that the air connection tubes are not bent, squashed or
crushed;
• Check that the chambers are inflated progressively and according
to the selected program;
• Make sure you set the pressure control correctly;
• Verify that the actual air outlet from the unit is present.
• Check for holes or damage in the applicators' air chambers. If no
anomalies are detected, switch off the device, switch it back on
and check that it is working properly again