Copper PP / M153
Copper PP
0.05 - 5 mg/L Cu
Bicinchoninate
Material
EN
Required material (partly optional):
Reagents
VARIO CU1 F10
VARIO CU1 F10
ValidCheck Copper 2 mg/l
Preparation
1.
Digestion is required for the determination of total copper.
2.
The pH value of the sample must be adjusted between 4 and 6 before analysis (with
potassium hydroxide solution or nitric acid).  Any resulting dilution must be taken
into account in the result.
Note: pH values above 6 can lead to Copper precipitation.
Notes
1.
Accuracy is not affected by undissolved powder.
12
Packaging Unit
Powder / 100 pc.
Powder /
1000 pc.
1 pc.
EN   Method Reference Book   1.0
M153
Cu
Part Number
530300
530303
48141525