Télécharger Imprimer la page

Balluff BTL7-A/C/E/G5-M-H/W-S32/S115/ Serie Notice D'utilisation page 43

Publicité

BTL7-A/C/E/G5 _ _ -M _ _ _ _ -H/W(8)-S32/S115/_ _ _
Magnetostrictive Linear Position Sensor – Rod Style
6
Calibration procedure (continued)
6.3.2
Adjusting
The detailed procedure for adjusting is
described on page 17 ff.
The factory set null point and end point is replaced by a
new start point and end point and the associated output
values can be adjusted. The start and end values can be
adjusted as you like up to the limits.
Adjustment is possible from serial number
120615000xxxxx xx onwards.
Steps
► Move magnet to the new start position.
► Read new start point by pressing the buttons.
► Set the new start value by pressing the buttons.
New start value
New null point
Fig. 6-4:
Adjust a new start value
www.balluff.com
► Move magnet to the new end position.
► Read new end point by pressing the buttons.
► Set the new end value by pressing the buttons.
New end value
Fig. 6-5:
Adjust a new end value
6.3.3
Reset
Before
Restoring the BTL to its factory settings.
After
The detailed procedure for the reset is
described on page 19.
Before
After
New end point
english
15

Publicité

loading

Ce manuel est également adapté pour:

Btl7-a/c/e/g5-m-h/w8-s32/s115/ serie