Télécharger Imprimer la page

Drive Medical Pacifica Elite Guide D'instructions page 14

Publicité

Les langues disponibles

Les langues disponibles

Recommended separation distances between portable and mobile RF communications
equipment and the Pacifica Elite Compressor Nebulizer 18070, 18071.
The Pacifica Elite Compressor Nebulizer is intended for use in an electromagnetic envi-
ronment in which radiated RF disturbances are controlled. The customer or the user of
the Pacifica Elite Compressor Nebulizer can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equip-
ment (transmitters) and the Pacifica Elite Compressor Nebulizer as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance din metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where Pis the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propa-
gation is affected by absorption and reflection from structures, objects and people.
Please dispose of this nebulizer in accordance with local, municipal,
state and federal law for electronic devices
Rev.6.03.3.17
14
.............................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL 18070, 18071
............................................................................................................
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √ P
0.12
0.38
1.20
3.79
12.00
m
80 MHz to 800 MHz
d = 1.2 √ P
0.12
0.38
1.20
3.79
12.00
800 MHz to 2,5 GHz
d = 2.3 √ P
0.23
0.73
2.30
7.27
23.00

Publicité

loading

Produits Connexes pour Drive Medical Pacifica Elite

Ce manuel est également adapté pour:

1807018071