R4870 - R4875
EN - Mounting instructions
Please read and carefully follow these instructions. Find the product number on the product label and ensure that these
assembly instructions match your product.
WARNING: Failure to comply with these instructions may result in accidents involving serious personal injury or
!
the product being damaged.
WARNING: If you suspect that any individual parts may be damaged, contact your Pressalit dealer or service
provider. Do not remove, attempt to repair or install parts yourself, unless directed to do so by Pressalit or its
agents.
WARNING: No modification of this equipment is allowed without express authorization from Pressalit.
WARNING: These instructions must be followed, and the product must be installed professionally by a qualified
person; e.g. a licensed plumber, in order to avoid personal injury or property damage while using the product.
WARNING: Pressalit cannot be held liable for injury to persons or damage to property caused by faulty mounting.
Installation location
This product is designed for indoor use in a wet environment, such as a bathroom or washroom. It may only be used in
an environment corresponding to the system's IP rating. Ensure that the installation location environment also complies
with the following operating parameters of the product:
Environmental factor
Temperature
Relative humidity
Atmospheric pressure
WARNING: Medical electrical equipment needs special precautions regarding electromagnetic compatibility
(EMC) and needs to be installed and put into service in accordance with the EMC information (provided in your
product's operating instructions). It is unlikely that the user will encounter problems with the product because of
inadequate electromagnetic compatibility. However, electromagnetic immunity is always relative and standards
anticipate environments of usage. Portable and mobile RF communications equipment can affect medical
electrical equipment. When the product is connected to the local mains power supply, we recommend the use of
electromagnetic shielding and similar preventative measures to avoid conducted emissions.
When choosing an installation location, ensure that there will be adequate space for movement of—and access to—the
product.
Assembly and installation
Mount the product against a wall in accordance with the illustrations in this assembly instruction. Connect the product
components—hand control unit, motor, control box and power cable—to each other as shown in figure 1. Ensure that
the connectors are pressed in as far as possible into their corresponding sockets.
Connection to the mains supply
There is a range of potential configurations to connect the product to the mains power supply. Figures 2, 3a, 3b and 3c
illustrate some alternatives. The product is supplied with a plug attached, but this may need to be replaced with an
IP-rated plug and socket combination appropriate to the location of the socket relative to moisture, or alternatively the
product may be hardwired to a mains supply.
WARNING: Ensure that the circuit on the mains supply side is isolated from power while connecting the product
to the mains supply.
WARNING: The product must be connected to the mains supply and a protective earth connection by a qualified
!
electrician in accordance with IEC 60601-1 and IEC 60364-7, as well as any locally applicable standards.
WARNING: Ensure that the voltage of the local mains supply to be connected is compliant with that stated on the
label on the product.
Operating range
5°C to 40°C
20% to 90% @ 30°C (condensation free)
700 to 1060 hPa
Pressalit Care
Telephone:
Pressalitvej 1
Telefax:
DK 8680 Ry
E-mail:
Danmark
Web:
2019-10-31 · Page 2/21
+45 87 88 87 88
+45 87 88 87 89
pressalit@pressalit.com