3.5 Scale selection
0 ÷ 4000 mg O
0 ÷ 999 mg O
0 ÷ 600 mg O
0 ÷ 250 mg O
0 ÷ 90 mg O
The scale is related to the expected value. The possibility of measuring diluted samples extends the range.
Results are shown as mg O
Wireless mode
Before the analysis set the scale on the RESPIROSoft™
3.6 Analysis duration and sampling time
Wireless mode
With RESPIROMETRIC Sensor an analysis can last from 5 to
120 days. The sampling time can be selected depending.
on the desired duration of the analysis, from 1 h to 2 days
Wireless mode
Before the analysis set the scale on the RESPIROSoft™
3.7 Alarms
AL1
Calibration error. Restart the analysis. If the problem persists contact the service.
AL2
Analysis start denied. Check on RESPIROSoft™.
AL3
Max. sensors q.ty reached. 50 analyses are already in progress. Wait the end.
AL4
One parameter or more are wrong
AL5
No PC response. Check the distance between the RESPIROMETRIC Sensor and the Wireless DataBox™.
AL6
The analysis is stopped from the PC
AL7
Unknown command result
AL8
Data sending failed. Restart the analysis. If the problem persists contact the service.
AL9
No PC response. Check if the RESPIROSoft™ is active.
AL10
No automatic measure is done
AL11
Wireless DataBox™ not found. Check if the Wireless DataBox™ is connected.
Wireless DataBox™ connection isn't strong enough
AL12
AL13
Analysis start is denied due to low battery level
bAt
Low battery (wireless mode OFF). Remove the batteries with the fingers and replace them with new ones paying
attention to the polarity indicated inside the instrument.
Err
Calibration error (wireless mode OFF).
---
The measured value overcomes the maximum value of the chosen scale.
4. Maintenance
RESPIROMETRIC Sensor warns about the required change of batteries when the unit is resetted and the charge is lower
than a minimum threshold. Use a coin lever in the rear part of RESPIROMETRIC Sensor and open it. Batteries are removed
using the fingers. Introduce two fresh batteries at the same time paying attention to the polarity. Battery replacement deletes
any stored value.
Scale
/l
2
/l
2
/l
2
/l
2
/l
2
PRE
BMP
/l (ppm) for BOD test and values in mbar for PRE scale.
2
Volume of sample
Refer to the method
Refer to the method
Non-wireless mode (after the reset)
Key SET (display on)
Key SET or START
Non-wireless mode
Respirometric Sensor will perform only 5-day test
with 24-hour sampling time for BOD analysis or
30 days using PRE scale.
6
25ml
100 ml
150 ml
250 ml
400 ml
The scale is changed
The scale is shown