7.6 Temp. Unit Default Setting
•
Use " " Up, " " Down to select "1" or "0" .
1 = °F
0 = °C
•
After adjusting the Temperature unit, push the "Enter" button, then press "ESC" to
save the data and return to the normal display.
7.7 pH Manual Temp. Setting
•
This procedure is only to adjust the manual temperature compensation value for pH
measurement.
•
The lower display will show:
M. TEMP. SET
^, v Enter:Y
•
Use " " Up, " " Down to select the desired manual Temp. compensation value.
•
Press "Enter" once, then press "ESC" to save the data and return to the normal
display.
7.8 Temp. Compensation Factor Setting
•
This procedure is only used for the Conductivity function.
•
Use " " Up, " " Down to select the Temp. Compensation Factor (% per °C) of the
measured solution.
•
After setting the desired value, press "Enter", then press "ESC" to save the data and
return to the normal display.
•
Temp. Compensation Factor is typically set to 2.0% per C degree.
7.9 CD (µS, mS), TDS (ppm) Default Setting
•
This procedure is only used for the Conductivity function.
•
Use " " Up, " " Down to select "1" or "0".
0 = µS, mS
1 = ppm
•
After adjusting the unit (µS/mS, ppm), press "Enter", then press "ESC" to save the
data and return to the normal display.
7.10 DO % Salt Compensation Value Setting
•
This procedure is only available for the DO function.
•
The lower display will show:
% Salt SET
^,v Enter:Y
36
AL 15