Measurement and instrumentation
Multimeter model UT60A
Réf :
291 033
ENGLISH
All manuals and user guides at all-guides.com
2 Specifications
2.1 General Specifications
•
Maximum Voltage between any Terminals and Grounding: 1000V rms
•
Fused Protection for µA mA input terminal: 0.5A, 250V fast type Glass
fuse, Ø 5x20 mm
•
Fused Protection for 10A input Terminal: 10A, 250V fast type Glass
fuse, Ø 5x20 mm
•
Maximum Display: Display: 3999
•
Measurement Speed: Updates 3 times /second
•
Temperature: Operating: 0°C – 40°C (32°F – 104°F)
Storage: -10°C – 50°C (14°F – 12°F)
•
Relative Humidity: < 75% à O°C – 30°C
•
Altitude Operating: 2000 m
Storage: 10000 m
•
Battery Type: One piece of 9V NEDA1604 or 6F22 or 006P.
•
Overloading: Display OL.
•
Auto and manual range selectable.
•
Dimensions (HxWxL) : 177 x 85 x 40 mm.
•
Weight: Approximate 300g (battery included).
•
Safety/Compliances: IEC61010 CAT III 1000V,
•
Certification: CE UL & CUL
2.2 Accuracy Specifications
•
Accuracy: ±(a% reading + b digits) guarantee for 1 year
•
Operating temperature: 23°C ± 5°C
•
Relative humidity: <75%
•
Temperature coefficient: 0.1 x (specified accuracy)/1°C
20
< 50% à 31°C – 40°C
CAT IV 600V overvoltage
Double insulation standard