All manuals and user guides at all-guides.com MANUAL DE INSTRUCCIONES OPERATING INSTRUCTIONS MANUEL D’INSTRUCTIONS MULTÍMETRO DIGITAL DIGITAL MULTIMETER MULTIMÈTRE DIGITAL COD.51255 ESPAÑOL ....... 2 ENGLISH ......18 FRANÇAIS ......34 GARANTIA/GUARANTEE/ GARANTIE ......49...
Page 2
All manuals and user guides at all-guides.com ESPAÑOL INFORMACION DE SEGURIDAD Este Medidor cumple con los estándares IEC61010: grado de contaminación 2, categoría de sobretensión (CAT I 1000V, CAT II 600V) y doble aislamiento. CAT.I: Nivel de señal, equipamiento especial o partes de equipamiento, telecomunicación, electrónica, etc..., con menores sobre voltajes transitorios que sobre voltajes CAT.II.
Page 3
All manuals and user guides at all-guides.com • No utilice o almacene el Medidor en un ambiente de alta temperatura, humedad, explosivo, inflamable y fuertes campos magnéticos. La capacidad del Medidor puede deteriorarse después de humedecerse. • Preste atención al aislamiento alrededor de los conectores. •...
Page 4
All manuals and user guides at all-guides.com ESTRUCTURA DEL MEDIDOR (VEA FIGURA 1) 1. Pantalla LCD 2. Botón retención de datos 3. Clavija transistor 4. Terminal de entrada COM 5. Otros terminales de entrada 6. Terminal de entrada mA 7. Terminal de entrada 20A/10A 8.
Page 5
All manuals and user guides at all-guides.com SIMBOLOS DE LA PANTALLA (VEA FIGURA 2) figura 2 Nº Símbolo Significado Retención de datos activado Indica lectura negativa The battery is low. Batería baja. Advertencia: Para evitar falsas lecturas, que puedan generar descargas eléctricas o daños personales, reemplace la bacteria tan pronto como el icono aparezca.
Page 6
All manuals and user guides at all-guides.com 2. Gire el selector giratorio hasta una posición de medida apropiada 3. Conecte los cables de prueba al objeto a medir. El valor medido será mostrado en la pantalla. negro rojo figura 3 Nota •...
Page 7
All manuals and user guides at all-guides.com C. MEDICION DE CORRIENTE DC (VEA FIGURA 4) Advertencia! Nunca intente medir la corriente de un circuito donde la tensión entre los terminales y tierra sea mayor que 60V DC o 30V rms. Si el fusible se quema durante la medición, el Medidor puede quedar dañado o el operador puede resultar herido.
Page 8
All manuals and user guides at all-guides.com D. MEDICION DE CORRIENTE AC (VEA FIGURA 4) Advertencia! Nunca intente medir la corriente de un circuito donde la tensión entre los terminales y tierra sea mayor que 60V DC o 30V rms. Si el fusible se quema durante la medición, el Medidor puede quedar dañado o el operador puede resultar herido.
Page 9
All manuals and user guides at all-guides.com negro rojo figura 5 Nota - Los cables de prueba pueden añadir un error de 0.1Ω a 0.3Ω a la medición de resistencia. Para obtener lecturas precisas en bajas resistencias, rango de 200Ω, los cables de prueba y memorice la lectura obtenida (llamada esta lectura como X).
Page 10
All manuals and user guides at all-guides.com figura 6 Nota • Para medir el condensador con polaridad, conecte el cable de prueba rojo en el ánodo y el negro en el cátodo. • Cuando el condensador medido sea corto o el valor esté sobrecargado, la pantalla mostrará “1”.
Page 11
All manuals and user guides at all-guides.com - Tensión de circuito abierto aproximadamente 2.8V. - Cuando la prueba de diodo haya sido completada, desconecte la conexión entre los cables de prueba y el circuito bajo prueba y retire los cables de los terminales de entrada. Para la prueba de continuidad, conecte el Medidor del siguiente modo: 1.
Page 12
All manuals and user guides at all-guides.com Modo Sleep Para preservar la vida de la batería, el Medidor se apaga de manera automática si no gira el selector o presiona ningún botón durante alrededor de 15 minutos. En ese momento, el Medidor consume en torno a 10 µA.
Page 13
All manuals and user guides at all-guides.com C. Corriente DC Precisión Rango Resolución Protección sobrecarga COD. 51255 200µA 0.1µA 0.315A. 250V 1µA ±(1%+3) fusible tipo rápido, 20mA 10µA Ø5x20mm 200mA 100µA ±(1.8%+3) 10A/20A 10mA ±(3%+5) Un-fused Observación: • En 10A: Para medición continua ≤ 10 segundos e intervalo no menor que 15 minutos. •...
Page 14
All manuals and user guides at all-guides.com Observación: • Tensión circuito abierto: A 200MΩ: aprox. 3V Otros rangos: ≤ 700mV F. Capacidad Precisión Rango Resolución Protección sobrecarga COD. 51255 200nF 0.1nF ±(4%+3) 250V AC 2µA 20µA 10nF Observación: • Señal: aprox. 400Hz, 40mV rms. G.
Page 15
All manuals and user guides at all-guides.com tornillo figura 8 figura 9 J. SUSTITUCION DE LA BATERIA (VEA FIGURA 8) Para evitar falsas lecturas, que puedan llevar a posibles descargas eléctricas o daños personales, sustituya la batería tan pronto como el indicador aparezca.
Page 16
All manuals and user guides at all-guides.com NOTAS IMPORTANTE! El fabricante no se responsabiliza de los daños o mal funcionamiento del aparato, en caso de que no se use correctamente o se haya utilizado para trabajos para los que no esté diseñado. De acuerdo con la Directiva de Residuos de Aparatos Eléctricos y Electrónicos (RAEE), estos deben ser recogidos y dispuestos por separado.
Page 17
All manuals and user guides at all-guides.com...
Page 18
All manuals and user guides at all-guides.com ENGLISH SAFETY INFORMATION This meter complies with the standards IEC61010: in pollution degree 2, overvoltage category (CAT. I 1000V, CAT. II 600V) and double insulation. CAT.I: signal level, special equipment or parts of equipment, telecommunication, electronic, etc., with smaller transient overvoltages than overvoltages CAT.
Page 19
All manuals and user guides at all-guides.com • Do not use or store the Meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after dampened. • When using the test leads, keep your fingers behind the finger guards. •...
Page 20
All manuals and user guides at all-guides.com THE METER STRUCTURE (SEE FIGURE 1) 1. LCD Display 2. Data Hold Button 3. Transistor Jack 4. COM Input Terminal 5. Other Input Terminals 6. mA Input Terminal 7. 20A/10A Input Terminal 8. Capacitance Jack 9.
Page 21
All manuals and user guides at all-guides.com DISPLAY SYMBOLS (SEE FIGURE 2) Nº Symbol Meaning Data hold is active Indicates negative reading The battery is low. Warning: To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears.
Page 22
All manuals and user guides at all-guides.com 2. Set the rotary switch to an appropriate measurement position in range. 3. Connect the test leads across with the object being measured. The measured value shows on the display. Note • If the value of voltage to be measured is unknown, use the maximum measurement position (1000V) and reduce the range step by step until a satisfactory reading is obtained.
Page 23
All manuals and user guides at all-guides.com C. DC CURRENT MEASUREMENT (SEE FIGURE 4) Warning! Never attempt an in-circuit current measurement where the voltage between terminals and ground is greater than 60V DC or 30V rms. If the fuse burns out during measurement, the Meter may be damaged or the operator himself may be hurt.
Page 24
All manuals and user guides at all-guides.com measurement. Use proper terminals, function, and range for the measurement. When the testing leads are connected to the current terminals, do not parallel them across any circuit. To measure current, do the following: 1.
Page 25
All manuals and user guides at all-guides.com Note - The test leads can add 0.1Ω to 0.3Ω of error to resistance measurement. To obtain precision readings in low-resistance measurement, that is the range of 200Ω, short-circuit the input terminals beforehand and record the reading obtained (called this reading as X). (X) Is the additional resistance from the test lead.
Page 26
All manuals and user guides at all-guides.com G. MEASURING DIODES & CONTINUITY Warning! Use the diode test to check diodes, transistors, and other semiconductor devices. The diode test sends a current through the semicondutor junction, then measure the voltage drop across the junction.
Page 27
All manuals and user guides at all-guides.com H. MEASURING TRANSISTOR (SEE FIGURE 7) To measure transistor, connect the Meter as follows: Set the rotary switch to hFE. 2. Insert the NPN or PNP type transistor to be tested into the transistor jack. 3.
Page 28
All manuals and user guides at all-guides.com A. DC Voltage Accuracy Range Resolution Overload Protection COD. 51255 200mV 100µV 250V DC or AC ms ±(0.5%+1) 10mV 1000V DC or 750V AC 200V 100mV 1000V ±(0.8%+2) Remark: • Input impedance: 10MΩ. B.
Page 29
All manuals and user guides at all-guides.com AC Current Accuracy Range Resolution Overload Protection COD. 51255 200µA 0.1µA 0.315A. 250V 1µA ±(1%+3) fast type fuse, 20mA 10µA Ø5x20mm 200mA 100µA ±(1.8%+3) 10A/20A 10mA ±(3%+5) Un-fused Remark: • At 10 A range: For continuous measurement ≤10 seconds and interval not less than 15 minutes.
Page 30
All manuals and user guides at all-guides.com F. Capacitance Accuracy Range Resolution Overload Protection COD. 51255 200nF 0.1nF ±(4%+3) 250V AC 2µA 20µA 10nF Remark: • Testing signal: approx. 400Hz, 40mV rms. Diodes and Continuity Test Input Function Range Resolution Remark Protection Open circuit voltage...
Page 31
All manuals and user guides at all-guides.com J. REPLACING THE BATTERY (SEE FIGURE 8) To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears. To replace the battery: 1.
Page 32
All manuals and user guides at all-guides.com NOTES IMPORTANT! The maker will not take responsibility for damage or malfunction as a result of the device being incorrectly used or, applied for a purpose for whith it was not intended. According to Waste Electrical and Electronic Equipment directive (WEEE), these ones must be collected and arranged separately.
Page 33
All manuals and user guides at all-guides.com...
Page 34
All manuals and user guides at all-guides.com FRANÇAIS INFORMATIONS DE SÉCURITÉ Ce mesureur est conforme avec les standards IEC61010 : degré de pollution 2, catégorie de surtension (CAT I 1000V, CAT II 600V) et double isolement. CAT.I : niveau de signal, équipement spécial ou parties d’équipement, télécommunication, électronique, etc., avec des inférieurs en voltages transitoires qu’en voltages CAT.II.
Page 35
All manuals and user guides at all-guides.com CA, on devra avoir une attention particulière au danger d’éventuelles décharges électriques. • Utilisez la fonction, la borne et le rang appropriés lors de toute mesure. • N’utilisez ni ne stockez le mesureur dans un environnement explosif, inflammable, à haute température, à...
Page 36
All manuals and user guides at all-guides.com STRUCTURE DU MESUREUR (VOIR FIGURE 1) 1. Écran LCD 2. Bouton rétention de données 3. Fiche transistor 4. Borne d’entrée COM 5. Autres bornes d’entrée 6. Borne d’entrée mA 7. Borne d’entrée 20A/10A 8.
Page 37
All manuals and user guides at all-guides.com SYMBOLES DE L’ÉCRAN (VOIR FIGURE 2) Nº Symbole Signification Rétention de données activée Lecture négative Batterie faible. Avertissement : Pour éviter de fausses lectures qui peuvent produire des décharges électriques ou des dommages personnels, remplacez la batterie dès que l’icône apparaît.
Page 38
All manuals and user guides at all-guides.com borne d’entrée COM. 2. Tournez le sélecteur jusqu’en position de mesure appropriée. 3. Connectez les câbles mis à l’épreuve afin de mesurer. 4. La valeur mesurée sera affichée sur l’écran. Noir Rouge Note •...
Page 39
All manuals and user guides at all-guides.com • Quand la mesure de tension en CA aura fini, débranchez la connexion entre les câbles mis à l’épreuve et le circuit sous épreuve. C. MESURE DE COURANT CC (VOIR FIGURE 4) Avertissement ! Ne tentez jamais de mesurer le courant d’un circuit dont la tension entre les bornes et la terre soit supérieure à...
Page 40
All manuals and user guides at all-guides.com D. MESURE DE COURANT CA (VOIR FIGURE 4) Avertissement ! Ne tentez jamais de mesurer le courant d’un circuit dont la tension entre les bornes et la terre soit supérieure à 60V CC o 30V rms. Si le fusible est brûlé lors de la mesure, le mesureur peut être endommagé...
Page 41
All manuals and user guides at all-guides.com Noir Rouge Note - Les câbles mis à l’épreuve peuvent ajouter une erreur de 0.1Ω à 0.3Ω à la mesure de résistance. Pour obtenir des lectures précises en résistances basses, rang de 200Ω, les câbles mis à l’épreuve et mémorisez la lecture obtenue (cette lecture étant appelée X).
Page 42
All manuals and user guides at all-guides.com Note • Pour mesurer le condensateur avec polarité, connectez le câble mis à l’épreuve rouge dans l’anode et le noir dans le cathode. • Si le condensateur mesuré est court ou si la valeur est surchargée, l’écran affichera ‘1’. •...
Page 43
All manuals and user guides at all-guides.com - Quand l’épreuve de diode soit complétée, débranchez la connexion entre les câbles mis à l’épreuve et le circuit sous épreuve et retirez les câbles des bornes d’entrée. Pour l’épreuve de continuité, connectez le mesureur de la manière suivante : 1.
Page 44
All manuals and user guides at all-guides.com Mode Sleep Pour préserver la vie de la batterie, le mesureur s’éteindra automatiquement si vous ne tournez pas le sélecteur ou n’appuyez aucun bouton pendant environ 15 minutes. En ce moment, le mesureur consomme environ 10 µA. Le mesureur peut être activé...
Page 45
All manuals and user guides at all-guides.com C. Courant CC Précision Rang Résolution Protection surcharge COD. 51255 200µA 0.1µA 0.315A. 250V 1µA ±(1%+3) fusible type rapide, 20mA 10µA Ø5x20 mm 200mA 100µA ±(1.8%+3) 10A/20A 10mA ±(3%+5) Un-fused Observation : • En 10A : pour une mesure continue de ≤...
Page 46
All manuals and user guides at all-guides.com Observation : • Tension circuit ouvert : À 200 MΩ : approx. 3 V Autres rangs : ≤ 700 mV F. Capacité Précision Rang Résolution Protection surcharge COD. 51255 200nF 0.1nF ±(4%+3) 250V AC 2µA 20µA 10nF...
Page 47
All manuals and user guides at all-guides.com Boulon J. REMPLACEMENT DE LA BATTERIE (VOIR FIGURE 8) Pour éviter des fausses lectures pouvant causer d’éventuelles décharges électriques ou des dommages personnels, remplacez la batterie dès que l’indicateur apparaît. Pour remplacer la batterie : 1.
Page 48
All manuals and user guides at all-guides.com...
Page 49
GARANTIA NO CUBRE AQUELLAS PIEZAS QUE POR SU USO NORMAL TIENEN UN DESGASTE. PARA OBTENER LA VALIDEZ DE LA GARANTIA , ES ABSOLUTAMENTE IMPRESCINDIBLE QUE COMPLETE Y REMITA ESTE DOCUMENTO A EGA MASTER , DENTRO DE LOS SIETE DIAS A PARTIR DE LA FECHA DE COMPRA.
Page 50
All manuals and user guides at all-guides.com...
Page 51
All manuals and user guides at all-guides.com...
Page 52
All manuals and user guides at all-guides.com C/ ZORROLLETA 11, POL. IND. JUNDIZ 01015 VITORIA, SPAIN P.O.B. APTDO. 5005 FAX. 34 - 945 290 141 TEL. 34 - 945 290 001 info@egamaster.com www.egamaster.com...