Fault
Polymerisation time
does not end automati-
cally
Polymerisation process
is not interrupted when
the cover is opened
Unsatisfactory polymeri-
sation results
Cause
LED circuit board (9) is
defective
Electronic module is
defective
Micro-switch (15) is
defective
Electronic module is
defective
Both micro-switches
(15) are defective
Polymerisation time is
too short
Material reacts to
fluctuating wavelengths
LED circuit board (9) is
soiled
LED circuit board is
damaged
Troubleshooting
The LED circuit board
must be examined by a
specialist, and replaced
if necessary
The electronics must
be examined by a
specialist, and replaced
if necessary
The micro-switch must
be checked by a spe-
cialist and replaced if
necessary
The electronics must
be examined by a
specialist, and replaced
if necessary
Deactivate the
device immediately.
The micro-switch must
be checked by a
specialist and replaced
if necessary
Extend the polymerisa-
tion time
Replace the material
Clean the LED circuit
board carefully
Examine the LED
circuit board, replacing
if necessary