Measurement and instrumentation
Multimeter model UT70A
Réf:
291 034
ENGLISH
All manuals and user guides at all-guides.com
•
Operating temperature: 5
0
0
5
C – 31
C / <50% to 31
•
Storage temperature: -10
•
Elevation: (operating) 2000 m, (storage) 10000 m.
2.2 Electricity
•
Accuracy: ±(a% reading + b digits) guarantee for 1 year
•
Operating temperature: 23°C ± 5°C
•
Relative humidity: <80%
DC voltage
Range
Resolution
200 mV
2 V
20 V
10 mV
200 V
100 mV
1000 V
Remarks: Input impedance approx. 10MΩ
AC voltage
Range
Resolution
200 mV
100 µV
2 V
1 mV
20 V
10 mV
200 V
100 mV
750 V
Remarks:
•
Input impedance: 10MΩ
•
Frequency reponse: 40 – 400 Hz.
•
Showing effective value of sine wave (mean value response).
DC current
Range
Resolution
20 µA
0.01 µA
2 mA
200 mA
100 µA
10 A
10 mA
Remarks: At « 10A Range » test no more than 10sec at interval of 15min.
AC current
Range
Resolution
20 µA
0.01 µA
2 mA
200 mA
100 µA
10 A
10 mA
13
0
0
0
C - 40
C (41
F - 104
0
0
C – 40
C.
0
0
0
C – 50
C (14
F – 122
Accuracy
10 µV
1 mV
± (0.5 % + 1)
1 V
± (0.8 % + 2)
Accuracy
± (1.2 % + 3)
± (0.8 % + 3)
1 V
± (1.2 % + 3)
Accuracy
± (0.8 % + 1)
1 µA
± (1.5 % + 1)
± (2 % + 5)
Accuracy
± (1 % + 3)
1 µA
± (1.8 % + 3)
± (3 % + 7)
0
F), humidity <80% to
0
F).
Input protection
500 Vrms
1000 VDC
750 VAC
Input protection
500 Vrms
1000 VDC
750 VAC
Input protection
F.0.5 A / 250 V
F.10 A / 250 V
Input protection
F.0.5 A / 250 V
F.10 A / 250 V