Accuracy (cont'd)
Accuracy (Cont'd)
All Models (cont'd)
Accuracy of Frequency Ranges
Display Range
10.00 Hz to 65.53 Hz
65.5 Hz to 655.3 Hz
0.655 kHz to 6.553 kHz
6.55 kHz to 50.00 kHz
Frequency—Logic Level Hz (mV Function)
Range
5.0 Hz to 6.553 Hz
6.55 Hz to 65.53 Hz
65.5 Hz to 655.3 Hz
0.655 kHz to 6.553 kHz
6.55 kHz to 65.53 kHz
65.5 kHz to 500.0 kHz
500.0 kHz to 655.3 kHz
0.655 MHz to 1.000 MHz
Non-contact Electric Field Detection (EF)
Typical Voltage
10 V to 36 V
23 V to 83 V
59 V to 165 V
124 V to 330 V
250 V to 1000 V
* Bar graph indication and tone are proportional to signal strength
Specifications
Display: LCD (6000) and 24-segment bar graph
Polarity: Automatic
Sampling Rate:
Numeric Display: 5 per second
Bar Graph Display: 40 per second
Temperature Coefficient: Nominal 0.15 x (specified accuracy) per °C below 18 °C or
above 28 °C
Automatic Power Off: After 34 minutes of inactivity.
To disable this feature, press SELECT while turning the meter on.
All manuals and user guides at all-guidesbox.com
Accuracy
± (0.03% + 0.03 Hz)
± (0.03% + 0.3 Hz)
± (0.03% + 0.003 kHz)
± (0.03% + 0.03 kHz)
Accuracy
± (0.03% + 0.002 Hz)
± (0.03% + 0.02 Hz)
± (0.03% + 0.2 Hz)
± (0.03% + 0.002 kHz)
± (0.03% + 0.02 kHz)
± (0.03% + 0.2 kHz)
± (0.03% + 0.2 kHz)
± (0.03% + 0.002 MHz)
Bar Graph Indication*
–
– –
– – –
– – – –
– – – – –
DM-200A • DM-210A • DM-510A
16
Sensitivity
(square wave)
3 V peak
5 V peak
Frequency Range
50 Hz to 60 Hz
19