5. Installation
• Place the incubator on a clean, level & stable surface. Position the appliance and allow it to stand for two or three hours (ten hours
in summer) before it is turned on for the first time. Do not position the unit next to any heat source e.g. boiler or direct sunlight.
• A clearance of approximately 13-15 cm must be left between the sides of the unit and any adjacent cabinets or walls to allow for
adequate air circulation. The unit is not designed for recessed installation.
• Remove the power supply from the white box inside the chamber and plug the wall end into the wall outlet. Plug the other end of
the power supply into the 2 pin inlet (top inlet) labelled DC onto the backside of the unit.
Internal Outlet
To use the internal outlet plug the included power cable into the bottom power inlet on the backside of the instrument, internal outlet
is now powered.
Note!
The voltage of the internal outlet is dependent on the wall voltage (if the wall voltage is 230 V, voltage of the
internal outlet is also 230V).