Data logging
Oxygen sensor
Ref : 482 039
- Measurement in water:
To make a measurement in water, change the calibration by pressing the orange button. The
arrows appear before the "Water" and "20 mg/L" O
the value of the displays is the same as for adjustment in air.
- Calibration in connected mode:
The oxygen sensor is automatically recognised by the onboard software.
An "Adjustment" tab is available to adjust the percentage of O
or in mg/L as required.
Remark:
- If you have already calibrated the probe in stand-alone mode, there is no need to repeat
the calibration using the software.
- The sensor display is inactive during use in connected mode.
3 Technical characteristics
Operating temperature range: +5/+45 °C
• Air Mode:
- Measuring range: 0 to 25 % O
- Accuracy: ± 1% of the maximum value full scale
- Number of decimal points in autonomous mode: 1
• Water Mode:
- Measuring range: 0 to 20 mg/L O
- Accuracy: ± 1% of the maximum value full scale
- Number of decimal points displayed in stand-alone mode: 1
ENGLISH
indications. The procedure for adjusting
2
2
2
7
. Simply enter the O
as a %
2
2