Data logging
Foxy force sensor
Ref : 482 047
2.3 Use in stand-alone or connected mode
2.3.1 In stand-alone mode
Adjustment:
Select the calibration then adjust zero:
A long press on the adjustment button: Segment
A short press on the adjustment button: Taring OK, segment
After adjusting zero, the value displayed is the instantaneous measured value.
2.3.2 When connecting to the computer
The sensor display goes out. The values measured are visible only on the computer
screen and the buttons are inactive.
The measuring principle is identical as for stand-alone mode, simply follow the
instructions on the computer screen.
3 Additional equipment
No additional equipment is required, except for the attachment systems (and the
console).
4 Technical characteristics
ENGLISH
Adjusting
zero
-
Calibrations: ± 5N, ±50N (displayed to 4 significant figures)
-
Accuracy: 0,1% of full scale
-
Resolution : 0,025 N
Calibration
flashes.
5
goes out.