The device is made up of a lead battery which cannot be accessed by outside. In order to replace it, consult the technical staff
authorised by the manufacturer.
USE ONLY THE RECOMMENDED BATTERIES FROM CA-MI. THE USE OF OTHER BATTERIES ARE NOT
RECOMMENDED AND INVOLVING THE CANCELLATION OF WARRANTY
In the event that the service personnel has to replace the internal battery, pay special attention to the polarity of the same
component. The + / - polarities are indicated directly on the battery.
If the overfill security system it's activated, don't proceede with the liquid aspiration. If the overfill security system doesn't work
there are two cases:
1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological filter who avoid the liquid
penetration inside the device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the device, in this case return the
device to CA-MI technical service.
CA-MI Srl will provide upon request electric diagrams, components list, descriptions, setting instructions and any other information
that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT
CA-MI TECHNICAL SERVICE. CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.
The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or
storage.
The working position must be such as to allow one to reach the control panel and to have a good view of the empty
indicator, the jar and the antibacterial filter.
It is recommended not to keep the device in your hands and / or to avoid prolonged contact with the body of apparatus.
WARNING: For proper use, place the aspirator on a flat, stable surface in order to have the full volume of use of the jar and
better efficiency of the overflow device.
Operation using with AC/DC switching adapter
Place the unit on a flat, horizontal surface.
Connect the short silicon tube, with antibacterial filter, to the suction connector. The other tube, with one end connected to
the antibacterial filter with the other end to jar's lid connector where has been fixed the red float. When the 90% of the
volume of the jar is reached there is the activation of the security float (the float close the aspiration connector on the jar) to
the avoid liquid penetration inside the device.
Connect the long silicon tube to the other jar's lid connector
Connect the other end of the long silicon tube to the probe plastic connector then connect the suction probe to it.
Connect the switching adapter to the device with the appropriate connector and insert the power cable plug to the power
socket. To start the treatment press the I switch to turn it on
Set the desired vaccum value (kPa / mmHg) with the appropriate vacuum regulator. Turning the handle clockwise increase
the vacuum value: these values can be read on the "vacuum indicator" instrument.
To suspend and / o r terminate the treatment, press the switch againg and pull the plug out from the power socket.
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of
the functionally vacuum) then rescrew the lid on the jar correctly.
To extract the accessories and start with cleaning.
WARNING: Ensure that the filter is fitted with the Arrows at the patient side.
The power supply cable plug is the element of separation from the electrical mains system: even if the units equipped
with a special on / off switch button, the power supply plug must be kept accessible once the device is in use so as to
allow a further method of disconnection from the mains supply system.