Before testing, verify the Receiver is functioning properly by placing
it near the powered Transmitter. With the Receiver set to Sensitivity
Level 8, a numerical reading of 99, all 10 signal strength bars lit up,
and a strong audible signal confirms the system is working. The
Receiver will produce a variable beep proportional to the sensitivity
value. The higher the signal strength, the faster it will beep.
NOTE: A continuous beep or solid-on LED indicates strongest
possible signal of 99 and the Receiver is saturated.
NOTE: If the Transmitter is powered up normally but the Receiver
detects a very low signal level, even when set to the highest
sensitivity level, the internal protection fuse in the Transmitter
should be checked, as it may be blown.
CONNECTING TEST LEADS
Connect test leads by inserting the black lead into the "COM" jack and
the red lead into the "+ 600V" jack. Do not test if leads are improperly
seated. Results could cause intermittent display readings. To ensure
proper connection, firmly press leads into the input jack completely.
NOTE: Test leads used must have a CAT III 600V safety rating.
CORRECT
INCORRECT
8