The petMAP gII is intended for use in the electromagnetic environment specified below. The
customer or user of the petMAP gII should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge
± 6 kV contact
(ESD)
± 8 kV air
IEC 61000-4-2
± 2 kV for power
Electrical fast
supply lines
transient/burst
IEC 61000-4-4
±1 kv for
input/output lines
± 1 kV differential
Surge
IEC 61000-4-5
± 2 kV common
Voltage dips,
short
>95% dip in 0.5
interruptions
and voltage
60% dip in 5 cycles
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Electromagnetic Immunity
IEC60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
mode
mode
cycle
30% dip for 25
cycles
>95% dip for 5 seconds
>95% dip for 5
seconds
3 A/m
Compliance Level
lines
±1 kv for input/output
lines
± 1 kV differential
mode
± 2 kV common mode
>95% dip in 0.5 cycle
60% dip in 5 cycles
30% dip for 25 cycles
3 A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a
typical commercial or hospital
environment.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
12