Ref : 453 076
4. Calibration
The probe must be calibrated in the following cases:
Operation Mode:
Notes:
4.1
4.1.1 Connect the probe to the adapter and the adapter to the ESAO
The adapter must be plugged in 15 minutes before beginning the calibration.
4.1.2 Move the selector on the adapter to the "air" setting.
4.1.3 Turn the slope button
4.1.4 Using the zero button, set the value displayed to 0.
Press the magnification button and use the zero button
The value displayed is slightly greater than 0 when you release the magnification button.
ENGLISH
Initial use
After changing the electrolyte
After changing the probe head
Whenever a precise measurement is required
 Ideally, the calibration should be performed at a temperature approximating that of
the experiment planned and in a moderately humid atmosphere (50 to 70% RH).
In practice, the ambient conditions of a laboratory or classroom are sufficient for a
satisfactory calibration.
 It should be noted that the level of CO
increase by 0.3% by the end of the period. It is therefore best to perform the
calibration in a well-ventilated room.
With your Primo/VTT or Visio interface
console.
Value displayed
Actual concentration
Response of the non-calibrated probe
in a classroom of twenty or so students can
2
to the highest setting (clockwise).
Value displayed
After
actuation
Actual concentration
Response after adjusting the slope button
13
Prior to
actuation
to display 0.0 at +/- 0.1.