value of the light intensity.
6.11 The optical fiber should be sterilized for 4 minutes with 134℃ and 2.0bar~2.3bar
(0.20MPa~0.23MPa) before each use.
6.12 The curing light is equipped with over-heat protection system. It can continuously work 200s,
For example, continuously operate the curing light for 10 times under 20s working mode (even
the curing light works less than 20s, it is counted as a full operation), then it will come into
over-heat protection status. And only after 2-minute sleep, it can restart working 200s
continuously.

7. Light intensity measurement

7.1 Connect the output plug of power adapter to the plug of DC5.0V in the pedestal.
7.2 Choose general mode and aim the optical fiber at the measurement point, press on / off
button, the present light intensity is displayed on the indicator of pedestal.
7. Light intensity measurement
7.1 Connect the output plug of power adapter to the plug of DC5.0V in the pedestal.
7.2 Choose general mode and aim the optical fiber at the measurement point, press on / off
button, the present light intensity is displayed on the indicator of pedestal.

8. Cleaning, Disinfection and Sterilization

The cleaning, disinfection and sterilization of optical fiber is as follow. Unless otherwise stated,
they will be hereinafter referred to as "products".
Warnings
The use of strong detergent and disinfectant (alkaline pH>9 or acid pH <5) will reduce the life
span of products. And in such cases, the manufacturer takes no responsibility. This device shall
not be exposed to high temperature above 138℃. Processing limit The products have been
designed for a large number of sterilization cycles.
The materials used in manufacture were selected accordingly. However with every renewed
preparation for use, thermal and chemical stresses will result in ageing of the products. The
maximum number of sterilizations for optical fiber is 500 times.
8.1 Initial processing
8.1.1 Processing principles
It is only possible to carry out effective sterilization after the completion of effective cleaning and
disinfection. Please ensure that, as part of your responsibility for the sterility of products during
use, only sufficiently validated equipment and product-specific procedures are used for
cleaning/disinfection and sterilization, and that the validated parameters are adhered to during
every cycle.
Please also observe the applicable legal requirements in your country as well as the hygiene
regulations of the hospital or clinic,especially with regard to the additional requirements for the
inactivation of prions.
8.1.2 Post-operative treatment
The post-operative treatment must be carried out immediately, no later than 30 minutes after the
completion of the operation. The steps are as follows:
the dirt on the surface of product with pure water (or distilled water/ deionized water);
2. Dry the product with a clean, soft cloth and place it in a clean tray.
Notes
a) The water used here must be pure water, distilled water or deionized water.
8.2 Preparation before cleaning