4.5 Size: 40×50×260mm
4.6 Net weight: 147g
4.7 Power consume: ≤8W
4.8 Protection type against electrical shock: ClassⅡ
4.9 Protection degree against electrical shock: Type B
4.10 Protection against harmful ingress of water or particular matter: ordinary equipment (IPX0)
4.11 Safety in the presence of flammable anesthetic mixture with air, oxygen or nitrous oxide: not
suitable under this condition
5. Installation and Demounting
5.1 Take off the red cap from the optical fiber, then insert the metal part into the front of main unit
(make sure to screw when installing, oblique insert should be prevented).
5.2 Fix the light hood on the bottom of the optical fiber.
5.3 Uninstall the LED, just reverse the procedure above.
5.4 Battery replacement method: open the battery cover of the main unit, take the battery out,
than disconnect the plug slightly. Connect the plug of the new battery correctly, put the new
battery in, and then fix the battery cover.
5.5 After using the machine or it is in need of charging, make the main unit headon inserted in
the pedestal and compress it to make sure the main unit and pedestal chucking. When there is
no need of charging, please pull out the pedestal Adapter or put main unit into the pedestal
inversely.
6. Operation
6.1 Press the mode button to set the working mode, the corresponding
indicator will on when a mode set.
6.1.1 Full-power: the blue light radiates in full power.
6.1.2 Ramping: the blue light power increases stronger continually, after 5 seconds reaches to
the highest power.
6.1.3 Pulse: the blue light works on the pulse condition.
6.2 Press the time button to set the solidifying time, 4 working time is available: 5, 10, 15, 20
seconds.
6.3 When operating, aim the optical fiber at the correct position, press the power button to start
or stop to emit of the blue light.
6.4 During operation, the blue light can be stopped by press the power button at any time.
6.5 A battery detect circuit is fixed inside the main unit, when low power is detected, the indicator
of the main unit will wink, please charge in time.
6.6 Connect the Adapter and the charge device correctly, put the main unit into the charge
device, the yellow indicator will enlightened, when charge finish, the green indicator will
enlightened and the yellow one will off. The green light is the power indicator of the charging
finish. If a battery error is detected, the indicator of the charge device will wink.
6.7 When operating finish, please clean the optical fiber with calico to avoid infecting the light
intensity.
6.8 This equipment will turn off automatically if don't any action within 2 minutes, turn it on by
press any button.
6.9 The effective light intensity of this equipment is much more higher than Halogen Lamp, The
solidified depth of the curing light composites resin for 10 seconds will not less than 4mm.
6.10 This equipment has the function of light intensity measuring. Connect the Adapter of the
light intensity measure device correctly, aim the top of optical fiber at the sensor, turn on the main
unit to emit blue light, the green indicator of the light intensity measure device will indicate the
18